|
|
|
Chapter V -
Radiation Control Measures for Analytical X-Ray Equipment
5.1 Administrative
Requirements for Use of Analytical X-Ray Equipment
5.1.1. Registration
Analytical x-ray equipment at ASU must be registered to a
single member of the faculty, academic professional, or staff
who bears responsibility for safe use of the equipment by all
other individuals. See Appendix
F.
5.1.2. Notification of Receipt of Equipment
The Office of Radiation Safety be notified within 30 working
days of the receipt of analytical x-ray equipment.
5.1.3. Radiation Surveys of Equipment
After initial installation, and upon completion of
alterations or maintenance, the Office of Radiation Safety must
conduct a survey for radiation leakage before the unit may be
placed in service. Surveys may be scheduled through the Office
of Radiation Safety. See Appendix
E.
5.1.4. Maintenance
Only properly trained maintenance personnel may install,
repair, or make other than routine modifications to the x-ray
generating apparatus and the tube housing apparatus complex.
5.1.5 Testing of Safety Devices
Safety interlocks and other safety devices will be tested
monthly by registrants. Records of these tests should be
maintained by the registrant for review by Office of Radiation
Safety or ARRA personnel.
5.1.6. Emergency Procedures
Written emergency procedures pertaining to radiation safety
shall be established for each x-ray producing apparatus by the
Radiation Safety Officer and posted in a conspicuous location.
These procedures list the telephone number(s) of the Radiation
Safety Officer and include the following actions to be taken in
case of a known, or suspected accident involving radiation
exposure:
- Notify the Radiation Safety Officer, and
- Arrange for medical examination.
5.1.7. Unattended Use of Equipment
Analytical x-ray equipment must not be left unattended while
the tube is energized unless:
- An interlock device is provided to prevent accidental
entry into the primary beam, and
- The stray radiation at any accessible point at a distance
of 10 inches from the tube housing containnment, as measured
with monitoring instrument appropriate for the energy range
generated, is no greater than 2mR per hour.
5.2 Equipment Requirements
5.2.1. Labeling
All analytical x-ray equipment must be labeled with a readily
discernible sign or signs bearing the radiation symbol and the
words:
- "CAUTION - HIGH INTENSITY X-RAY BEAM", or words
having a similar intent, on the x-ray source housing; and
- "CAUTION - RADIATION - THIS EQUIPMENT PRODUCES
RADIATION WHEN ENERGIZED", or words having similar intent,
near any switch that energizes an x-ray tube.
5.2.2. Visual Indicators
- The primary on-off switch for each tube must include a
visual indication of the tube status, in the form of a warning
light on the control console and a warning light on the tube
housing, wired so that the x-ray tube cannot be energized if
the warning light fails. Such lights must operate at all times
when the tube is energized and must light at no other times.
- A shutter status (open or closed) indication must be
provided in the area adjacent to the tube head so that the
position of the shutter is readily discernible.
5.2.3. Interlocks
- An interlock device which prevents entry of limbs,
fingers, hands, wrists, etc., into the primary beam or causes
the primary beam to be shut off, must be utilized, unless
otherwise approved by the Radiation Safety Committee.
- It must not be possible to resume operation without
resetting the beam "ON" switch at the control panel.
5.2.4. Beam Shutters
On open-beam configurations, each port on the radiation
source housing must be equipped with a shutter that cannot be
opened unless a collimator or a coupling has been connected to
the port.
5.3. Facility Requirements
5.3.1. Dedicated Work Space
- X-ray diffraction and spectrographic equipment must be
placed in a room separate from other work area, unless
otherwise approved by the Radiation Safety Committee.
- Access to rooms containing analytical x-ray equipment must
be secured to prevent unauthorized use of equipment.
- All installations must display warning signs on the
outside of all entrances to the room. Warning signs will be
supplied and posted by the Office of Radiation Safety staff.
5.4. Radiation Limits
5.4.1. General
The local components of an analytical x-ray system must be
located and arranged, and must include sufficient shielding or
access control such that no radiation levels in excess of 80 mR
per hour will exist in areas accessible to the fingers, hands,
or forearms, or in excess of 5 mR per hour will exist in areas
accessible to the whole body, lens of the eyes, blood forming
organs, or gonads. When analytical x-ray equipment is placed in
a room utilized by personnel other than operators of the x-ray
producing equipment, the limits for these radiation levels will
be reduced by a factor of twenty.
5.4.2. Other Radiation Limits
- Each x-ray tube housing must be so constructed that with
all shutters closed the leakage radiation measured at a
distance of 5 cm from its surface is not capable of producing
a dose in excess of 2.5 mrem in one hour at any specified tube
rating.
- Each x-ray generator must be supplied with a protective
cabinet which limits leakage radiation measured at a distance
of 5 cm from its surface such that it is not capable of
producing a dose in excess of 0.25 mrem in one hour.
- Any apparatus utilized in beam alignment procedures must
be designed in such a way that excessive radiation will not
strike the operator. Particular attention should be given to
viewing devices, in order to ascertain that lenses and other
transparent components attenuate the beam to an acceptable
level.
5.5 Requirements of Users
5.5.1. Training and Other Responsibilities
No individual will be permitted to act as an operator of a
particular instrument until such individual has received an
acceptable amount of training in the correct use of specific
equipment and in radiation safety and is approved by the
Radiation Safety Officer. Operators are responsible for:
- Keeping radiation exposure to themselves and others as low
as is practical.
- Being familiar with safety procedures as they apply to
each instrument.
- Wearing of personnel monitoring devices, if required.
- Notifying the Radiation Safety Officer of known or
sususpected excessive radiation exposures.
5.5.2. Bypassing or Altering Interlocks and Safety
Devices
If, for any reason, it is necessary to temporarily
intentionally alter safety devices, such as bypassing interlocks
or removing a shield, such action shall be:
- Specified in writing and posted near the x-ray tube
housing so that other persons will know the existing status of
the machine.
- Terminated as soon as possible.
5.5.3. Other
- Personnel must not expose any part of their body to the
primary beam.
- Unused tube head ports must be secured in the closed
position. These must be checked prior use when the machine has
been left
unattended.
|
 |
 | |